How to change from Celsius to Fahrenheit, Change the battery and get the last reading on your Vicks Digital Thermometer. V966/V912
How to Change from Fahrenheit to Celsius.
Start with the thermometer OFF. Hold down the center button until the display shows Lo. Then release the button. Press the button again to move between Fahrenheit (F) or Celsius (C), to choose the desired scale. Once the “C” or “F” starts to blink, push the button again to turn off the thermometer.
Turn on the thermometer and the display will be in Fahrenheit (F) or Celsius (C) based on the scale you set in the previous steps. Your thermometer will default to Fahrenheit (F)when you replace your battery.
How to change the battery.
The battery is depleted and needs replacing when the “▼” symbol appears and flashes in the center top of the display. Using a coin, unlock the battery cover and carefully remove the battery.
For models: V912, V912US or V912USV1, V966, V966US, V966USV1- Replace with a 3V, CR1225 silver oxide battery.
For models: V912USV2, V912USV3, V912BBUSV2, V912BBUSV3, V966USV2, V966USV3 - Replace with a 3V, CR1632 silver oxide battery .
Insert the new battery in the positive (+) side facing the battery cover. Replace the battery cover using a coin and tighten. Make sure the cover is completely secure after replacing the battery.
How to get the last reading.
To get the last reading on your thermometer, press the power button and continue to hold for 2 seconds. The last measured temperature will be shown in the display with the letter “M” under the degree sign for 3 seconds. The unit will then shut off.
For more information visit VicksThermometers.com
This works.. you only have to do it right: after seeing "Lo " click immediately again . Don't wait until the "F" or "C" starts blinking.
THANK YOU I WAS SO CONFUSED
2032 is a Universal battery. Use of a 2032 battery was my #1 search criteria for oral thermometers. I almost passed over this one until I read the comments that it will accept a 2032. The video is doing a great disservice to this thermometer.
This never works for me. It only works to change from the default F to C. If it’s on C simply remove the battery for 30 seconds and reinstall it. It will default back to F. Make sure you write down the C reading because the memory recall will clear when the battery is removed.
Can't open the battery cover.
V916C-Aus model batteries are CR1220 (3V)
Thank you. That was helpful
Battery compartment would not open. Pried it open and it was hard. I just tossed it.
You made this way harder then it had to be
CR2032 batteries can be used in place of CR1225 batteries
Just bought one. Battery cover is stuck it does not fall off easily like it shows in the video. Waste of money
I keep on trying to use my thermometer to measure the heat coming from my hand, but it keeps on switching to the LO button, how can I fix this?
Do you have to add a degree still if used under the arm?
It doesn't work with my thermometer. the model is model V966F-24. Do you have any other way?
what does the M mean? mine doesnt make any noises
What does M mean? Mine doesn't display a body just the letter M
I can’t get the back cover open. Is there a trick to that?
stuck on Celsius. won't change to Fahrenheit
It does not change the scale, when I press the button it just turn off!