Case 10: PCI Manual - Stent loss

Sdílet
Vložit
  • čas přidán 31. 01. 2018
  • After failed attempts to deliver a stent through an in-stent restenotic lesion of the proximal right coronary artery, the stent came off the balloon while trying to retrieve it into the guide catheter. Part of the stent was inside and part was outside the guide catheter. A balloon was advanced next to the lost stent and inflated, followed by withdrawal of both the guide catheter, the lost stent, and the guidewire en bloc. The lesion was rewired, predilated and successfully stented with an excellent final result.
  • Jak na to + styl

Komentáře • 5

  • @aksuyeksavascelebi
    @aksuyeksavascelebi Před 2 měsíci

    hi, I want to share some part of that video (chart which is at 2:15) with your permission? ı will use that in my presentation in Europcr 2024.

  • @areenal-taie6836
    @areenal-taie6836 Před 3 lety +1

    Can we advance a 2nd guidewire before retrieving the 1st guidewire in blog along with the inflated balloon and undeployed stent ? Thanks a lot

  • @sutenly1
    @sutenly1 Před 6 lety

    can removed stent from artery?

  • @salahuddinsalahuddin3210
    @salahuddinsalahuddin3210 Před 4 lety +1

    Thanks for the nice case presentation.
    Is there a special reason dir using iFR first and FFR at the end?

    • @manosbrilakis
      @manosbrilakis  Před 4 lety

      No special reason - with iFR of 1.0 the likelihood of having an FFR of .80 or lower is essentially zero!