    Nice presentation. I'm curious about the dry out step. Does it increase the risk of membrane cracking?

    • @anneweeks12
      @anneweeks12  Před rokem

      Not at all, this parameter has been validated thoroughly and the membrane is robust enough to withstand this step.

  • @kalpamehta
    @kalpamehta Před 2 lety

    Thanks Michel for a great presentation. What are the differences in validation between the Plus pump and Oasis? Also if we add a Oasis adaptor head to Plus pump, does it push us to revalidate the whole system again?

    • @tcser7
      @tcser7 Před 2 lety

      HI Kalpa. No, you wouldn't need to re-validate the entire system because you won't be using the weighing function or any of the other functions inside the Milliflex Plus. You can run a performance test to ensure the pump works with the new head and supplement that with our validation guide.

  • @sirskaven7624
    @sirskaven7624 Před 2 lety

    What's the lever on the adapter for? And is there a proper way to remove the filter to prevent air pockets?

    • @anneweeks12
      @anneweeks12  Před 2 lety

      Hello Sir, the lever is simply to release the vacuum for easy removal of the membrane device from the filtration base. This is only necessary when using the adapter on a manifold, not the Oasis pump. There is no risk of air pockets, as the system is designed for full contact upon transfer. Hope this helps!