CT Components (Pictorial Explanation)

Sdílet
Vložit
  • čas přidán 15. 08. 2022
  • CT components are the important pieces of a CT scanner including:
    The x-ray tube,
    Pre-patient Bowtie Filter,
    X-ray collimator,
    High Voltage power supply,
    Heat exchanger,
    CT detector,
    Data Acquisition system,
    Slip Ring
    Table,
    Gantry Covers,
    Gantry Controls,
    Gantry Display,
    System cabinet,
    Liquid Tube Cooling,
    Reconstruction Computer
    Some of the key concepts covered in this video are the:
    CT Components
    CT scan components
    ct scan
    CT parts
    components of ct scan
    radiologic technologist
    ct scanner without cover
    ct scan video
    ct scan machine
    ct scanner
    computed tomography (ct)
    computed tomography physics
    ct scanner how it works
    cts components
    ct scan parts and functions
    ct gantry components
    ct scanner installation
    ct scanner machine
    ct scanner spinning without cover
    computed tomography basic principles
    computed tomography scan
  • Věda a technologie

Komentáře • 16

  • @talkingtreeswithprofessort8372

    Alright, alright, another great video Dr. Nett ;)

  • @user-lh5zn9vm2e
    @user-lh5zn9vm2e Před 5 měsíci

    Good Day Doctor, In the CT scan series description for a brain CT, it says "Base 0.75 H31s," while for different series, it says "Base 4.5 H60s." On internet dose report pages, I see "head" written. Is it normal for my dose report to say "base" instead of "head"? It seems like the cerebellum is not included in the imaging values because it says "base" instead of "head." On the internet, I see scans for brain CT described as "base+cere," but my scan with Siemens 16 Sensation only mentions "base." Are the details in the Series Description the same as in the Dose Report, and if one says "base," should the other also say "base"? Or have only the base region values been included in my imaging values? Therefore, is it not accurate to say total DLP is 927? Thank you in advance for your response.

    • @HowRadiologyWorks
      @HowRadiologyWorks  Před 5 měsíci +1

      Sorry we don’t comment on specifics of medical exams. The different series have different reconstruction kernels. The dose report shows include all doses.

    • @user-lh5zn9vm2e
      @user-lh5zn9vm2e Před 5 měsíci

      Last sentence is help me a lot doctor, so It means Even though the Study Description and dose reports mention "base" instead of "head" the values are for the entire scan.
      Thank you

  • @kimwagemans2081
    @kimwagemans2081 Před 9 dny

    verry clear explaind thank you, can you explain the pre detectror collimator as well.

  • @minhmo6918
    @minhmo6918 Před 3 měsíci

    🎯 Key Takeaways for quick navigation:
    00:00 🔄 *CT scanners consist of a rotating gantry that houses various components crucial for image acquisition.*
    01:37 📏 *Different types of filters, such as flat filters and bowtie filters, are used in CT scanners to adjust x-ray spectra and reduce patient dose.*
    04:36 🎯 *Collimators control the shape and size of the x-ray beam in CT scanning, with collimation specifications typically set at the iso center of the system.*
    Made with HARPA AI

  • @rg942
    @rg942 Před 3 měsíci

    on modern CT scanners does each detector have a separate collimator? thanks

    • @HowRadiologyWorks
      @HowRadiologyWorks  Před 3 měsíci +1

      There are a variety of designs but usually the xy collimator is for every detector column and the z collimator is usually a little less frequent

  • @panamazus
    @panamazus Před 2 měsíci

    What parameters control the bow tie filter used? My assumption is the scanned FOV…is this accurate?

  • @arceem870
    @arceem870 Před rokem

    So it lowers the patient dose, right sir? Can it be also attached to the detector?

    • @arceem870
      @arceem870 Před rokem

      I'm referring to the bow tie filter by the way

    • @HowRadiologyWorks
      @HowRadiologyWorks  Před rokem +1

      @@arceem870 yes the bow tie filter does reduce dose by filtering out X-rays on the way to the patient so that towards the outside of the patient there will be less X-rays. We would not want to put on detector thoughts as those X-rays have been through the patient so we don’t want to filter. The exception is an anti scatter grid as we do want to remove scattered X-rays before they hit the detector

    • @arceem870
      @arceem870 Před rokem

      @@HowRadiologyWorks Oh! I see, thank you so much sir. That cleared my confusions