HOW TO troubleshoot conditioning and drift issues with a coulometric Karl Fischer cell

Sdílet
Vložit
  • čas přidán 24. 05. 2021
  • In this video, the team at Metrohm UK & Ireland talk you through what to do if the coulometric Karl Fisher cell is not conditioning down, and the drift value remains high or is unstable.
  • Věda a technologie

Komentáře • 3

  • @jaredjones2443
    @jaredjones2443 Před rokem

    What do you do if everything is absolutely fine but it still won't condition? The engineer you keep sending is useless btw

    • @MetrohmTV
      @MetrohmTV  Před rokem

      Hi Jared
      Thank you for your comment. We would like to follow up on this with you and your local Metrohm subsidiary. Could you kindly send an email to info@metrohm.com with your contact details and your country of residence?
      Thank you and best regards
      Metrohm International Headquarters

    • @SashaBuzko
      @SashaBuzko Před rokem +1

      Likely, you have figured it out by now, but I'll suggest a possible cause in case someone else comes across this video. You could have a bad batch of the reagent. If the ratio imidazole/iodine is wrong, it will auto-generate water and the drift will always be over the limit. That's if you are totally sure that all parts of the titrator are clean and dry.