RT Clinic: Debunking an Oxygen Tubing Myth

Sdílet
Vložit
  • čas přidán 13. 08. 2023
  • Debunking the myth that the oxygen concentrator flow needs increased for more tubing. This is specific to home oxygen.
    Thanks for watching the RT Clinic!
    Please SUBSCRIBE to the channel! These videos are made for clinicians, students, and patients. I encourage comments, questions, and conversation below.
    --RT CLINIC BADGE REFERENCE--
    ETSY LINK
    rtclinicshop.etsy.com
    Thanks for watching the RT Clinic.
    Please consider following us:
    / thertclinic
    / thertclinic
    --Check out our "Light in the Darkness" Merch--
    rt-clinic.creator-spring.com/
    ----AUDIO CREDITS----
    Intro Music: Render One - Explasm
    If you would like your product demonstrated by the RT Clinic, please reach out to me via email. Jimmy.McKanna@gmail.com

Komentáře • 13

  • @sylvisterling8782
    @sylvisterling8782 Před měsícem

    I wish that tubing was MODIFIABLE as to length and connectors! My client has a 35 foot apartment range of motion... and until recdntly, I could ONLY find 25 foot and 50 foot tubing! The 25 footer meant that the client could not reach the front door because the distance from the concentrator was too far... and the 50 foot tubing was constantly underfoot, under wheelchair wheels, getting split, kinked, stepped on, tripped over... what a nightmare! But the only ones I can find have built-on connectors at each end. I FINALLY found a 40 foot tubing... HALLELUJAH!... but there's still more tubing than we need!

  • @shanshan8113
    @shanshan8113 Před 9 měsíci +1

    Can you make a video demonstrating the proper setup when transporting a patient in the hospital using oxygen?

  • @tymesho
    @tymesho Před měsícem

    Home O2 users need only a 4 dollar "monometer" for simple/daily checks on a tank regulator to liquid O2 dispersal.

  • @alejandrobonilla1141
    @alejandrobonilla1141 Před 10 měsíci +1

    Thank you so much for this video 🙌🙂

  • @michaelott310
    @michaelott310 Před 11 měsíci +1

    Why didn't you test an actual O2 concentrator? I've actually tried different lengths of tubing with patients on O2 concentrators. I didn't have a tester for the O2 flowrate. Instead, I used a pulse oximeter and noted the changes as I increased the tubing length, while keeping the flow at 2 lpm. The 7', 14 ' and 21' lengths showed consistent pulse oximetry of 95%. At 28' it dropped to 90%. Pts. were unaware of the changes. Their breathing pattern, rate and depth remained constant. I've also noticed desaturations O2 when you attach a concentrator to a BIPAP at the same flow rate as a nasal cannula with the same length of tubing. Perhaps a difference in flowmeters? Compensated vs uncompensated? I'm just telling you what I observed when I tried it on actual patients. I really love your videos. They're a great teaching tool for when I do RT in-services with nurses.

    • @rtclinic
      @rtclinic  Před 11 měsíci

      That's interesting. If I can get my hands on the testing unit, I'll do it with a concentrator next time. Maybe a CZcams short.

  • @EmmettGlenn
    @EmmettGlenn Před 5 měsíci

    What do you say to the patient who says, "I don't care what your devices and data say! I absolutely feel more breathless when I'm hooked up to a 50-foot cannula, with NO extensions. I breathe much better when I use my 7-foot cannula!"?
    My suggestion is to keep this video in your pocket so that the patient can watch it.
    This is very generous of you to educate those of us who see common sense logic in the myths we swear are true. Yes, you debunked the oxygen tubing myth.

    • @rtclinic
      @rtclinic  Před 4 měsíci +1

      If a patient said that, I would suggest they just use a 7ft cannula. 😊👍

    • @EmmettGlenn
      @EmmettGlenn Před 4 měsíci

      @@rtclinic And how would you suggest he use his O2 concentrator? Lug it around on a leash?
      (Uh-oh. My emojis aren't working. You'll never know I meant what I said in the spirit of fun and good-naturedness?)

  • @brockblack8765
    @brockblack8765 Před 11 měsíci +1

    Debunked indeed

  • @MohamedSaber007
    @MohamedSaber007 Před 11 měsíci

    Thanks for the great demonstration but I have a couple of questions about nebulization.
    -When using a nebulizer mask setup that is hooked up to the wall oxygen source does that require humidification or not? Should I replace the Christmas tree with a water cup or is it unnecessary?
    -When the medication is over does the mask with medication hub setup hooked up to an o2 source work as an oxygen delivery device or the patient isn't getting any breathable oxygen from that mask? because I've seen some co-workers do that and when I asked them they explained that a nebulizer mask setup could double as an oxygen delivery device but I don't think that's how it works, so is that true or is another myth that needs to be debunked?

    • @rtclinic
      @rtclinic  Před 11 měsíci +1

      Great questions.
      --It is unnecessary to use humidity with a nebulizer set up. The nasal cannula humidity should be removed from the flowmeter before attaching the nebulizer.
      -- A nebulizer mask will deliver oxygen after the medication is complete, but it is a variable amount of oxygen. I wouldn't recommend doing it because it is poor patient care. If oxygen is delivered, the clinician should have a general idea of the FiO2.
      This might be a good experiment though. 🤔 !

    • @MohamedSaber007
      @MohamedSaber007 Před 11 měsíci

      @@rtclinic Thanks a lot for clearing that up.