YOKOGAWA Oxygen Analyzer ZR402G

Sdílet
Vložit
  • čas přidán 6. 06. 2017
  • Transmitter Checking

Komentáře • 30

  • @shafiqkharal3351
    @shafiqkharal3351 Před 7 lety +1

    Wonderful it's easy and informative for Analyzer person

  • @kaminibrahma9831
    @kaminibrahma9831 Před 3 lety +1

    You could have added mounting of accessories and field calibration with actual O2 concentration cylinders.

  • @dorjnamjigsanjaasuren337

    hi. In screen the of analyzer displaying cell temperature is 334. error 1. Can you please tell me what do i have to do?

  • @instrumentpower6380
    @instrumentpower6380 Před 2 lety +1

    Sir please make field calibration vedio

  • @razzaqulhaider440
    @razzaqulhaider440 Před rokem +1

    Dear sir, you not clear, what is the value set at decode board. Can I used veritable resistor alternat decode board

    • @instrumentationandsolution5013
      @instrumentationandsolution5013  Před rokem

      PT1000 RTD used for cold junction compensation, u need to put atmospheric temperature ( the resistance value around 1110 to 1120 ohms)

  • @wed789
    @wed789 Před 5 lety

    Dear sir,my yokogawa o2 analyzer value is going high upto 2-3% give me suggestion for rectify it.

    • @instrumentationandsolution5013
      @instrumentationandsolution5013  Před 5 lety

      Anurag Awadhiya.. 1st you have to do proper calibration..and check calibration gas expire date .. while calibration check mV according to chart. mV vs O2 chart ,

  • @user-po9dn3ot5c
    @user-po9dn3ot5c Před 5 měsíci +1

    क्या कोई बता सकता है, दिल्ली की हवा में कितने प्रतिशत आक्सिजन है ?

    • @instrumentationandsolution5013
      @instrumentationandsolution5013  Před 5 měsíci

      Multi gas analyzer lagana hai Bhai ... Wo bohut costly hai... Pollution is going to increase everywhere 😔😔😔😔....

  • @sirajuddinidris8134
    @sirajuddinidris8134 Před 3 lety +1

    how to reset sir,? I have problem sir, error 02, heater temp. please help

  • @tunguyenvan5925
    @tunguyenvan5925 Před 5 lety

    PASS WORK OF IT

  • @rajeshkumar-vm6og
    @rajeshkumar-vm6og Před 6 lety

    Sir error 2- element heater reset krne pr bhi nhi ja rha h ...
    Please provide me suggestion for solution it..

    • @instrumentationandsolution5013
      @instrumentationandsolution5013  Před 6 lety

      Remove TC cable in Transmitter side and apply around 33 mV/750 deg C to Transmitter side. if u r error is reset then check TC is OK or NOT. if TC is OK then check at heater wire you should get around AC voltage 60 to 120VAC. and u can check heater COIL resistance etc.

    • @rajeshkumar-vm6og
      @rajeshkumar-vm6og Před 6 lety

      Anil Kumar kalla
      Sir heater coil, RTD , and thermocouple is ok..
      Sir normaly 338°c temp. Show to kr rha pr temp. Increase nhi ho rha h aur na hi auto me heater me supply ja rha h temp. Rise hone k liye..21% O2 show ho rha h display pr...

    • @instrumentationandsolution5013
      @instrumentationandsolution5013  Před 6 lety

      So simulate the transmitter. then you can find out the problem

  • @Indianghumakkad1993
    @Indianghumakkad1993 Před 3 lety

    Sir how to solve alarm 10 , display shows C.J temp

    • @instrumentationandsolution5013
      @instrumentationandsolution5013  Před 3 lety

      plz check CJ @ PT-1000 resistance once ......and Inside temperature of the transmitter also... CJ range is -25 to 155 deg C, if value exceeds then CJ alarm will generate....

    • @Indianghumakkad1993
      @Indianghumakkad1993 Před 3 lety +1

      @@instrumentationandsolution5013 Error 2 heater temp how to solve

    • @instrumentationandsolution5013
      @instrumentationandsolution5013  Před 3 lety

      @@Indianghumakkad1993 Please check heater Thermocouple and heater coil once, either TC or heater having problem,

  • @rajeshkumar-vm6og
    @rajeshkumar-vm6og Před 6 lety

    Sir apka no. Mil skta h?
    Ek problem h apse bt krna h agr ho ske to de please give me...

  • @powerplantci7287
    @powerplantci7287 Před 4 lety

    हिंदी में बता देता भाई

  • @pankajpatel662
    @pankajpatel662 Před 3 lety

    Send yokogawa service no