MRI - QUALITY ASSURANCE

Sdílet
Vložit
  • čas přidán 22. 07. 2024
  • I’ve been thinking about this topic for a while, and I also find this topic very important to exclude faulty equipment. I will show you both on old and newer Siemens software how you easily can do this by your own. Artifacts can arise not only from parameter settings, but also from coils and other equipment. Don't miss this opportunity to enhance your knowledge
    Remember to check out and support my PATREON here: / bacmri
    Thanks for watching, following and support.
    #S_MAGNETIC_VLOG #MRI #SIEMENSHEALTHINEERS #QA #QUALITY_ASSURANCE #MEDICAL_IMAGING #MRI_TECHNOLOGIST #MEDICINE #MRI_RADIOGRAPHER #MRIRADIOGRAPHER #RADIOLOGY #PATREON
    References:
    Phlanr - Phase (Vlog No Copyright Music): rb.gy/g0phc0

Komentáře • 21

  • @meoluoi7466
    @meoluoi7466 Před 22 dny +1

    tuyệt vời quá anh ơi
    chủ đề này hay lắm anh

  • @aneetaedwards3775
    @aneetaedwards3775 Před dnem

    hi bac while using T2 haste fatsat with low TR and flip angle will decrease the image contrast,

  • @user-hq3nz9ql9g
    @user-hq3nz9ql9g Před 23 dny +1

    Nice video. Lots of useful information. Where can I find this program so I can do practise on my computer?

  • @aneetaedwards3775
    @aneetaedwards3775 Před 19 dny

    hi bac i got a posting in ge mri, and i had tough time in DWI in lower limbs, iam uable to get proper homogenous images i really struggle a lot if can help, please.

  • @wodiagazian9527
    @wodiagazian9527 Před 6 dny +1

    Hi Bac, one of your followers here, how do you plug in table position on Siemens when scanning Cervical to lumbar without recentering manually? I do that on GE to where I plug in I for inferior average 360 then it gets it 90% of the time?

    • @Nguyen_MRI
      @Nguyen_MRI  Před 6 dny

      Hi You mean automatically coil selection?

    • @wodiagazian9527
      @wodiagazian9527 Před 4 dny +1

      @@Nguyen_MRI no sir, let’s say you start brain at -4 location then what do you plug in for lumbar? It is on the tab where you have climate comtrol

    • @Nguyen_MRI
      @Nguyen_MRI  Před 4 dny

      @@wodiagazian9527 czcams.com/video/LDbRncgNwxM/video.htmlsi=VO8BzOEYjNtXGGNw
      Use iso and move the fov it should automatically choose correct elements. If not go into the system and coils. There is an option for auto coil select.

  • @SHARONSHAJI248
    @SHARONSHAJI248 Před 22 dny

    Can You upload a video about MRI quality control?

  • @frankiechankf1
    @frankiechankf1 Před 23 dny +1

    done by mri radiographer

  • @abrabr6720
    @abrabr6720 Před 23 dny +1

    How if you have same artifact “noise” however the result of the coil qa is ok? thanks in advance.

    • @Nguyen_MRI
      @Nguyen_MRI  Před 23 dny

      @@abrabr6720 exclude coil faulty

    • @abrabr6720
      @abrabr6720 Před 23 dny +1

      @@Nguyen_MRI ?
      Do you think is related to protocol issue, since it is more obvious on the stir sequence?

    • @Nguyen_MRI
      @Nguyen_MRI  Před 23 dny

      It's coil faulty. Was easier to see on FS image.

  • @Totoro00007
    @Totoro00007 Před 16 dny +1

    MRI radiographer

  • @swargofficial4118
    @swargofficial4118 Před 7 dny

    Sir please Do Video sempra 1.5t i Have need Sequence Specifi But No One Give Me and Help Me

  • @local93
    @local93 Před 23 dny +1

    So we don’t analyse qa results. We only get the answer it’s ok or not ok? Seems too easy 😅

    • @Nguyen_MRI
      @Nguyen_MRI  Před 23 dny

      not ok means you know it's coil faulty and go on from there with Siemens if having service deal

  • @swargofficial4118
    @swargofficial4118 Před 7 dny

    Neck Cor t1 artifice coming How Can Remove artifice