Memmert Waterbath Training - Operating the Basic Controller

Sdílet
Vložit
  • čas přidán 11. 02. 2020
  • www.lasec.com/

Komentáře • 12

  • @mounaghodbane9732
    @mounaghodbane9732 Před rokem +2

    Bonjour , j'ai un bain marie memmert modele wnb10 , qui marche très bien mais dernièrement la température cesse d'augmenter au delà de 20 degrés es ce qu'il ya une solution merci d'avance7

  • @zhiezhie4306
    @zhiezhie4306 Před 5 měsíci +1

  • @MiguelFloresV.
    @MiguelFloresV. Před 4 měsíci

    How can I calibrate the temperature?

  • @benis-one8773
    @benis-one8773 Před 3 lety

    hello, i have a little problem, in the display showing a symbol of "temperature protection (TB)", any solution??

    • @LasecGroup
      @LasecGroup  Před 3 lety

      Hi Beni, All water baths are fitted with a mechanical temperature limiter (TB) Protection Class 1 to DIN 12880.
      If the electronic control unit should fail during operation and the fixed factory-set maximum temperature is exceeded by approx. 30°C the temperature limiter switches off the heating permanently as a final protective measure. The symbol lights up continuously as a warning.
      In addition to its function as overtemperature protection, the TB also operates as low-level protection,
      i.e. the heating is switched off permanently if the liquid drops below a certain level. As a warning, the symbol lights up continuously.
      Please see page 13 of this document, www.memmert.com/index.php?eID=dumpFile&t=f&f=5315&token=8a24842996f805be3740965c13a73a63dd431a41, for possible solutions to fix. It is recommended to contact your supplier directly for further customer support so as to not damage the equipment.

    • @benis-one8773
      @benis-one8773 Před 3 lety +1

      @@LasecGroup okay, thanks for your advice, i'll try it first

  • @hazwanahmad2273
    @hazwanahmad2273 Před rokem

    Hello. I’m trying to set the temperature to 80c but the warning ⚠️ sign keeps beeping and there is an E-L indicator on the screen. Any advice? Thank you

    • @LasecGroup
      @LasecGroup  Před rokem

      Hi Hazwan, E - L advises that communication to the power unit has been interrupted/disrupted; Try plugging it into another plug point and see if the same error message occurs? If the same error message appears, we recommend contacting your supplier directly to have the equipment serviced to avoid any damage.
      You may also look at pages 19 and 23 of this document, www.memmert.com/index.php?eID=dumpFile&t=f&f=5318&token=a5754c94ae408da30fe5b23c41ea693148e8c08e, for possible explanations for the issue.

    • @kakysamir9681
      @kakysamir9681 Před 7 měsíci

      Liquid level too low

  • @yousramhm7419
    @yousramhm7419 Před rokem

    Good morning
    What kind of water should I use for memmert waterbath

    • @LasecGroup
      @LasecGroup  Před rokem

      Hi Yousra, Operation should be with non-flammable thermostating liquids only! The tray may suffer damage as a result of corrosion and pitting, causing the temperature control
      fluid to infiltrate the heating system. Only ever use demineralised water with a conductance level of 5-10 microsiemens and a pH value between 5 and 7. The tray may suffer damage if ultrapure water or DI water with an electrical conductance level of below 5 microsiemens is used. Only ever use pretreated water with a conductance level of 5-10 microsiemens.
      Please see page 8 of this document, www.memmert.com/index.php?eID=dumpFile&t=f&f=5318&token=a5754c94ae408da30fe5b23c41ea693148e8c08e.

  • @regenerador8593
    @regenerador8593 Před rokem +1

    Video malo