"Range Confusion"? Which DMMs Have This Problem?

Sdílet
Vložit
  • čas přidán 16. 06. 2024
  • Some multimeters give erroneous readings when the selected current range does not match the designated jacks the leads should be in. I went through some of the DMMs I have and let's find out which ones have this problem.
    00:00 Problem demonstration with Zotek ZT-702S
    02:40 Kaiweets ST600Y and ANENG AN009
    04:03 HoldPeak HP-770D, BK 2709B
    05:59 HT118E, HT118A
    07:40 ANENG AN870, UNI-T UT61E
    08:40 ET829, Hantek 2D72
    09:58 OWON HD200 series, conclusions
  • Věda a technologie

Komentáře • 30

  • @boots7859
    @boots7859 Před 11 měsíci +16

    Why you only have 20k subs is beyond me. All your videos are well done, informative and detailed enough without a lot of off-topic meandering or algo-driven drama. Thanks.

  • @tanmayvadhera4250
    @tanmayvadhera4250 Před 11 měsíci +21

    so the conclusion is ... buy blue multimeters : )

  • @JeffinTD
    @JeffinTD Před 4 měsíci +2

    Your channel is absolutely excellent. Thank you for all the time and effort that went into these.

  • @newoldbrain
    @newoldbrain Před 7 měsíci +2

    You have the most oddball collection of DMMs I've ever seen!
    Just checked my Fluke 87V and EEVblog GW121, as I never thought of this problem.
    They both have a single position for A and mA, so the range is dictated by where the probe is inserted.
    Both also have a separate uA selector position and will beep if the probe is in the A socket, showing "LEAD" on the display.

  •  Před 3 měsíci +4

    You are the savior of the third world country engineers and hobbyist, as a low income person wasting money on Chinese instruments is a must, letting people know wich one is the best among the worst is the best thing, you are doing God's work over here.

  • @SlaVoy
    @SlaVoy Před 11 měsíci +9

    Nice video. I will include such a test in my upcoming multimeter reviews. Owon continues to pleasantly surprise me, and over time, I'm growing to like it even more (despite the highly non-intuitive signal adjustment on the oscilloscope).
    Zoyi unfortunately is the worst in this regard. If I saw over 2A instead of the expected 20mA during some tests with my face close to the PCB, I would immediately jump up from my chair :D

  • @jspencerg
    @jspencerg Před 11 měsíci +4

    Thanks for the chapter marks. I just checked the 61e+. There are so many more choices since I bought that meter.
    Seems I just need to keep agreement between dial selection and input plug used. Meter works fine for me.

  • @andrimastirta4736
    @andrimastirta4736 Před 2 měsíci +1

    Amazing, you gave very detail explanation ❤👍

  • @wilsonluilui
    @wilsonluilui Před 11 měsíci +3

    Thx kerry wong it give out a great point of current measure,And for the series of test current be careful of never turn the selection switch of multimeter when power still on,it must burned the meter ,for my meter dead now😅.

  • @anthonyheak3479
    @anthonyheak3479 Před 11 měsíci +1

    Nice and clear explanation. Thanks

  • @Pootycat8359
    @Pootycat8359 Před 11 měsíci +4

    5:20 My Fluke 87III does that too, but it's MUCH too "polite." It squeaks softly, no doubt because it doesn't want to startle you.

  • @PeregrineBF
    @PeregrineBF Před 11 měsíci +4

    This can be an annoyance, but it's something you should be double-checking before measuring current because if you get it wrong the other way around you'll blow a fuse (by Murphy's law will not have spares on hand). If the meter doesn't warn you you're on the wrong range *before* you try to take readings it's not a very good meter. If it gives the wrong results while beeping loudly at you, that's OK IMO. If it can use a single set of jacks for μA through A, that's even better (though that's usually reserved for bench meters).

  • @VitaliyHTC
    @VitaliyHTC Před 11 měsíci +3

    Have you ever opened multimeter? Did you saw shunt inside for 10A range?
    If you plugging probes in wrong holes, what you expect?
    RTFM!

  • @JohnUsp
    @JohnUsp Před 9 měsíci +3

    Interesting. I always use Amps, because my Amps have so much resolution to measure even less than 1mA and can support up to 20A for 30 seconds.

  • @barrybogart5436
    @barrybogart5436 Před 11 měsíci +3

    I think that both you and Dave Jones are asking too much of manufacturers, and not enough of users. I have many of the meters you have both tested, including the 702S and am never confused. But then I have been doing electronics since 1955, using a Simpson 260 and a Triplett 310. But wait-both of these models are STILL made almost unchanged. I guess they just make them for us codgers who read manuals. YOU do the most thorough reviews and I appreciate it and vote with my purchases. I trust you a lot more than Dave.

    • @KerryWongBlog
      @KerryWongBlog  Před 11 měsíci +1

      Thanks Barry. Just wanted to point out for unsuspecting users. I haven't made these mistakes either but I thought it was still worth pointing out as after all there are meters out there with all these potential issues well thought out.

  • @user-ep7lh6tp1d
    @user-ep7lh6tp1d Před měsícem +1

    Thanks for your

  • @rokiedecentra9656
    @rokiedecentra9656 Před 11 měsíci +1

    things that never come to mind when assuming you can trust the reading displayed....

  • @miguelc_seeker
    @miguelc_seeker Před 11 měsíci +1

    Greetings, I have three UNIT UT61+, UT89XD, UT191T multimeters and a GOLDSTAR DM-311 from 1996 +/- ....and I performed those tests measuring current, which you did and none of them gave me a bad result, they all measured perfectly. use loads from 5mA to 40mA in DC and AC... they all measured well, when changing the place of the probe and changing the dial as you did... without problems 👍
    saludos, poseo tres multimetros UNIT UT61+, UT89XD, UT191T y un GOLDSTAR DM-311del año 1996 +/- ....y realice esas pruebas midiendo corriente, que tu hiciste y en ninguno me dio mal, todos median perfecto. utilice cargas desde 5mA hasta 40mA en DC y en AC.... todos midieron bien, al cambiar las sonda de lugar y cambiar el dial como tu lo hiciste.... sin problemas 👍

  • @khalifaalzaabi8230
    @khalifaalzaabi8230 Před 11 měsíci

    Hello my friend, if you can review the scale UNI-T UT117C

  • @JanioRibeiroSilva
    @JanioRibeiroSilva Před 11 měsíci +3

    I have the Uni-t UT181a and Agilent U1252b I did the same tests none of them let me measure the scale with the terminal connected to the ampere scale wrongly👍

  • @antoineblaise8365
    @antoineblaise8365 Před 9 měsíci

    Hello everyone, would you please to tell me what is the best multimeter?

  • @hitorsh1t
    @hitorsh1t Před 3 měsíci

    Is your source dc constant current or pulsed

  • @muppetpaster
    @muppetpaster Před 7 měsíci

    4:05 Here the meter is not confused....The user is....Meter does what it should.....Just put it in the right range...Seriously....Those are NOT incorrect readings but incorrect SETTINGS.

  • @shaunmorrissey7313
    @shaunmorrissey7313 Před 11 měsíci +5

    You really are making a mountain out of a mole hill this time, you clearly state you get an incorrect reading when in the WRONG setting.

    • @rokiedecentra9656
      @rokiedecentra9656 Před 11 měsíci +4

      not really, this is still a good point to bring up as sometimes you might make such a mistake and get wrong readings. Knowing about such issues will ensure the mistake doesn't happen

    • @johncoops6897
      @johncoops6897 Před 11 měsíci +3

      Agreed. Even worse.... ALL of my meters give the incorrect Volts and Ohms measurements when I have the probes in the Amperage jacks. Go figure!

    • @boots7859
      @boots7859 Před 11 měsíci +2

      Give me a break, anyone who does any work with DMM's will make this mistake more than once. Sometimes you get so far ahead in your troubleshooting or checking that you completely forget.
      Yes, thats never happened to you, but for the rest of us mortals, it does happen.

    • @orlandoalarcon4158
      @orlandoalarcon4158 Před 27 dny

      Solo puedo regañarle que muestre valores en amperios cuando está en el rango de miliamperios eso sí llama un poco la atención