Programat CS - Chapter 4 - Calibration & Dehumidification

Sdílet
Vložit
  • čas přidán 25. 01. 2011
  • Instructional Video - Chapter 4

Komentáře • 16

  • @zedfreyparco6132
    @zedfreyparco6132 Před 7 lety +1

    how to to trouble shoot the error 800 of programat p510, i already check the vacuum pump and clean it, cunduct dehumidification program

    • @Ivoclar-global
      @Ivoclar-global  Před 7 lety

      Dear Zedfrey Parco
      Please perform a vacuum test and let us know how many mbar the furnace reaches and shows after 4:50 min. on the display.
      Please do not hesitate to contact me directly: serviceline@ivoclarvivadent.com
      Best regards
      ServiceLine, Ivoclar Vivadent

  • @raulochoa1136
    @raulochoa1136 Před 3 lety +1

    I have a p300 g2 I did maintenance on it and now it only turns on the green led, no display comes out and no solution turns on, thanks

    • @Ivoclar-global
      @Ivoclar-global  Před 3 lety

      Dear Raul Ochoa
      I kindly ask you to contact one of our local distributor directly.
      www.ivoclarvivadent.com/en/contact/distributors
      Best regards
      ServiceLine, Ivoclar Vivadent

  • @raulochoa1136
    @raulochoa1136 Před 3 lety +1

    tengo un p300 g2 le hize mantenimiento y ahora solo prende el led verde no sale display y no prende alguna solucion gracias

    • @Ivoclar-global
      @Ivoclar-global  Před 3 lety

      Estimado Sr. Raul Ochoa
      Muchas gracias por su mensaje.
      Le rogamos entre en contacto con nuestro distribuidor.
      www.ivoclarvivadent.com/en/contact/distributors
      Cordiales saludos
      ServiceLine, Ivoclar Vivadent

  • @Wally17Feb
    @Wally17Feb Před 6 lety

    Please help!
    I'm getting error 802,,
    (No vacuum increase could be measured. Check the following points: Is the firing chamber tight (nocontamination on the sealing surfaces)? Is the vacuum hose connected? Is the vacuum pump connected?Is the fuse F1 o.k.?)

    • @Ivoclar-global
      @Ivoclar-global  Před 6 lety +1

      Dear Wael Najem
      If you checked all of that, then we kindly ask you to contact one of our local distributor.
      Best regards
      ServiceLine, Ivoclar Vivadent

  • @DogecoinMode
    @DogecoinMode Před 3 lety

    Can I use ATK2 for a second time?

    • @Ivoclar-global
      @Ivoclar-global  Před 3 lety

      Dear Modestas Bliudzius,
      No, the ATK 2 calibration can only be used once.
      Best regards,
      ServiceLine, Ivoclar Vivadent

  • @drtanya4ut
    @drtanya4ut Před 4 lety

    How do you know if dehumidification is completed?

    • @Ivoclar-global
      @Ivoclar-global  Před 3 lety

      Dear drtanya4ut,
      As soon you start the program you see a timer and the time is running back, if the timer is on zero, the program is finished.
      Best regards,
      ServiceLine, Ivoclar Vivadent

  • @darkdark4154
    @darkdark4154 Před 3 lety

    Please help error 1300

    • @Ivoclar-global
      @Ivoclar-global  Před 3 lety

      Dear Dark Dark,
      Error 1300 is a hint. It means, that the last calibration was made a long time ago.
      If you have any further questions please contact us directly.
      serviceline@ivoclarvivadent.com
      Best regards
      ServiceLine, Ivoclar Vivadent

  • @mahefuzaalam4170
    @mahefuzaalam4170 Před 5 lety

    Sari aap hindi language mai video dalo na

  • @user-er8kr1ld6z
    @user-er8kr1ld6z Před 11 měsíci

    eror 800,это что за ошибка?