Infors-ht Minifors 2

Sdílet
Vložit
  • čas přidán 28. 08. 2024

Komentáře • 5

  • @user-jt7hn7cj3o
    @user-jt7hn7cj3o Před 5 měsíci

    hello!, I'm trying to export the data from the device to an USB directly but it's not happening. Do you know which would be the problem, should something be activated previously?

  • @milibitschn
    @milibitschn Před 5 lety

    I have a question regarding the temperature control. It seems that at some point the system is heating the vessel although it is at the desired temperature. The device will not start cooling until we disconnect and reconnect the cooling water. Is it a software issue?

  • @fahmidazamanirin
    @fahmidazamanirin Před 6 lety

    I have a question about autoclave. For autoclaving, we separate every parts from the vessel holder or we don't need to seperate. Just we autoclave whole apparatus.

    • @invitrotechnologies-lifesc7969
      @invitrotechnologies-lifesc7969  Před 6 lety

      Hi Fahmida, there’s no need to separate the parts from the vessel holder. The vessel, vessel holder, reagent bottles and pump heads are all autoclaved together as a closed system, ensuring sterility.

    • @prydazb6073
      @prydazb6073 Před 2 lety

      @@invitrotechnologies-lifesc7969 Thank you for the guidance. Do we leave the pH probe and all other probes inside the vessle before autoclaving? Would you please share links for tutorial videos for setting up the infors? Thank you very much in advance.