CT Protocols
0:00 Intro
0:08 CT Requisition Grading Process
0:48 What's In A Protocol
1:12 Sample CT Head Protocol
1:32 Patient Preparation
2:04 Scan Instructions, Positioning
2:33 Z-axis coverage
2:41 Injection Parameters
4:01 Technical Parameters
4:34 Dose and Image Quality
6:28 Detector Configuration
6:51 Reconstructions, Rendering and Image Processing
8:03 Additional Comments
8:15 Protocol Development
Script and Audio-Only: drive.google.com/drive/folders/1HV19j4sgHiFpBZF8at1uRBtr4soHevvx?usp=sharing
zhlédnutí: 5 945

Video

CT Advanced and Emerging Applications
zhlédnutí 2,6KPřed 2 lety
0:00 Intro 0:08 CT Angiography (CTA) 0:27 Bolus Monitoring / Triggered Studies 1:50 Cardiac CT, ECG Gating 2:08 Prospective Gating 2:35 Retrospective Gating 3:07 Calcium Scoring 3:34 Virtual Endoscopy (Colonoscopy, Bronchoscopy, Angioscopy) 4:11 Dual Energy CT (DECT) 5:58 CT Simulation (Radiation Therapy Planning) 7:00 CT-Guided Interventional Radiology Procedures 7:33 Cone Beam CT (CBCT) 7:54 ...
CT Quality Control
zhlédnutí 11KPřed 2 lety
0:00 Intro 0:19 QC Role of All Technologists (Warm-up, Air Calibrations) 1:05 QC Tests 1:26 Water Phantom 1:36 CT Number Accuracy 1:53 Cross-Field Uniformity 2:38 Noise 3:12 CT Number Linearity 3:56 CT Slice Thickness (CT Tomographic Section Thickness) 4:28 Spatial Resolution 4:51 Modulation Transfer Function 5:06 Contrast Resolution (CT Low Contrast Detectability) 5:53 Patient Dose 6:40 Image ...
CT Dose
zhlédnutí 11KPřed 2 lety
0:00 Intro 0:07 Absorbed Dose 0:13 Equivalent Dose 0:27 Effective Dose 0:41 CT Dose Index (CTDI) 2:04 Dose-Length Product (DLP) 2:22 Dose and Image Quality 3:50 Technical Factors and Dose 4:15 Automatic mA modulation 5:22 In-Field Bismuth Shielding 6:15 Filtration, Bowtie Filters 7:00 Out-of-Field Lead Shielding Script and Audio-Only: drive.google.com/drive/folders/1HV19j4sgHiFpBZF8at1uRBtr4soH...
CT Image Reconstruction
zhlédnutí 44KPřed 2 lety
0:00 Intro 0:17 Back Projection 1:20 Filtered Back Projection 1:34 Filters 1:49 Sharpening Filter 2:30 Smoothing Filter 3:05 Sharp vs Smooth 4:27 Iterative Reconstruction 5:52 Multi Planar Reformatting (MPR) 7:06 Rendering Modes: Averaging, MIP, MinIP 8:45 3D Volume and Surface Rendering Script and Audio-Only: drive.google.com/drive/folders/1HV19j4sgHiFpBZF8at1uRBtr4soHevvx?usp=sharing
CT Contrast Media
zhlédnutí 16KPřed 2 lety
0:00 Intro 0:14 Iodinated Contrast Media 0:30 Routes of Administration 1:04 Oral Contrast 1:24 Intravenous (IV) Contrast 1:54 Central Venous Access Devices (CVADs) 2:07 Risk Factors and Consent 3:10 Scan Delay and Phases 3:57 Power Injection 4:41 Normal Side-Effects 5:04 Adverse Reactions 6:09 Kidney Function, Contrast-Induced Nephropathy 6:50 Estimated Glomerular Filtration Rate (eGFR) 7:14 Ex...
CT Image Quality
zhlédnutí 10KPřed 2 lety
0:00 Noise 0:30 Signal-to-Noise Ratio 0:54 Resolution 1:03 Spatial Resolution (High-Contrast Resolution) 1:31 Contrast Resolution (Low-Contrast Resolution) 1:56 Temporal Resolution 2:31 Improving Spatial Resolution 4:24 Improving Contrast Resolution 5:35 Summary on Image Quality and Dose Script and Audio-Only: drive.google.com/drive/folders/1HV19j4sgHiFpBZF8at1uRBtr4soHevvx?usp=sharing
CT Housfield Units and Windowing
zhlédnutí 11KPřed 2 lety
0:10 Image Production, Reconstruction 0:33 Linear Attenuation Coefficient 1:45 Back Projection 2:02 CT Numbers, Hounsfield Units 2:49 Calculating CT Numbers 3:37 Windowing 4:08 Window Width 6:09 Window Level 6:54 Combining Window Width and Window Level Script and Audio-Only: drive.google.com/drive/folders/1HV19j4sgHiFpBZF8at1uRBtr4soHevvx?usp=sharing
CT Pitch and Field of View
zhlédnutí 19KPřed 2 lety
0:03 Helical Pitch 1:17 Low Pitch 1:35 High Pitch 2:00 Low Pitch (Oversampling) 2:41 High Pitch (Undersampling) 3:40 Field of View 3:49 Scan Field of View (SFOV) 4:29 Display Field of View (DFOV) 5:33 Field of View Summary Script and Audio-Only: drive.google.com/drive/folders/1HV19j4sgHiFpBZF8at1uRBtr4soHevvx?usp=sharing
CT Basics: Major Components
zhlédnutí 21KPřed 3 lety
0:06 Comparison: CT to conventional radiography; pixels vs voxels. 0:52 1st and 2nd generation CT scanners 1:24 3rd generation (modern) scanners 1:47 Multi-row detectors 2:11 External components: Generator, Gantry, Table, Z-axis, console. 3:03 Internal Components: Tube, Detector, Data acquisition system 3:18 Slip Ring Technology 3:53 Helical and Axial Scan modes 4:47 Internal Components: Beam O...

Komentáře

  • @samuelvernon2397
    @samuelvernon2397 Před 2 měsíci

    U should go step by step with bone and lung... especially when writing out number line

  • @ziafezi4047
    @ziafezi4047 Před 3 měsíci

    As a radiologist in the making this helped me tonnes! Thankyou so much!!!

  • @champsrt21
    @champsrt21 Před 4 měsíci

    amazing!

  • @Ayat_Rad_Tales
    @Ayat_Rad_Tales Před 5 měsíci

    Very informative , thanks for sharing

  • @BlessUpDiHerbs
    @BlessUpDiHerbs Před 5 měsíci

    do the ct boards make you do math questions about this type of thing or is it more general familiarity?

  • @brunoknox295
    @brunoknox295 Před 6 měsíci

    Thats just not true in 80 and more % of the cades that increasing the pitch is decreasing the dose ore the other way. Show me a scaner where this is the case in dose modulated scans (there is no one). Thats only true with scaners during not dose modulated scans with GE and Canon. Because there are working with ma and not mas like siemens or Philips.

  • @juhasznandor6349
    @juhasznandor6349 Před 6 měsíci

    Simple & easy to understand. Thank you!

  • @danielking2604
    @danielking2604 Před 6 měsíci

    Thank u sir. U are a life saver. U made the concept easier to understand.

  • @wildanHayadinhanya
    @wildanHayadinhanya Před 6 měsíci

    Nice job .and thank you

  • @expositosmith7739
    @expositosmith7739 Před 7 měsíci

    As before, clear, complete and simple. What is not easy.... Thank´s for your job.!

  • @expositosmith7739
    @expositosmith7739 Před 7 měsíci

    Clear and great explanation. Thank´s for your job !

  • @anaghanp7227
    @anaghanp7227 Před 8 měsíci

    Very informative Thank you sir

  • @anaghanp7227
    @anaghanp7227 Před 8 měsíci

    Very informative Thank you sir

  • @ramjoshi4019
    @ramjoshi4019 Před 8 měsíci

    nice explanation sir...

  • @user-fv9tb9no6h
    @user-fv9tb9no6h Před 9 měsíci

    i feel like im watching blues clues again. thank you this is amazing. 😭

  • @user-fv9tb9no6h
    @user-fv9tb9no6h Před 9 měsíci

    amazing video thank you! For those studying with MOSBY pg 91 and a practice question say the generator is now house IN the gantry.

  • @user-fv9tb9no6h
    @user-fv9tb9no6h Před 9 měsíci

    omg this saved me THANK YOU!!

  • @thibautmodrzyk6215
    @thibautmodrzyk6215 Před 11 měsíci

    Very well explained !!

  • @thibautmodrzyk6215
    @thibautmodrzyk6215 Před 11 měsíci

    Very well explained !!

  • @alexalex78
    @alexalex78 Před 11 měsíci

    Great video. Who assigns the the WW and L? Is it the same for all vendors/machines?

  • @aditimishra5853
    @aditimishra5853 Před 11 měsíci

    Wonderful explanation

  • @alexalex78
    @alexalex78 Před 11 měsíci

    Thank you very much. Very informative.

  • @TheNettforce
    @TheNettforce Před rokem

    Thanks for the explanations, keep up the good work. One clarification is that at pitch greater than 1 there is typically not a gap and the data is not undersampled. There is less than a full scan of data but reconstruction can still be very accurate in this scenario. I know some books show this as a gap but this is a misnomer. Above pitch 1.37 or so is where data starts to be missing depending on DFOV.

  • @thecreator1127
    @thecreator1127 Před rokem

    Thank you so much :)

  • @thecreator1127
    @thecreator1127 Před rokem

    Thank You So Much

  • @psays
    @psays Před rokem

    Very helpful thank you!

  • @Minha-li4fk
    @Minha-li4fk Před rokem

    Thank you! your explanation is perfect !

  • @___Laura
    @___Laura Před rokem

    This video is amazing at explaining the scanner. Thank you so much!

  • @gurudattthakkar
    @gurudattthakkar Před rokem

    Very nice explanation

  • @blackheavy6781
    @blackheavy6781 Před rokem

    Thank you

  • @dianang5722
    @dianang5722 Před rokem

    Can I please clarify - increasing pitch will increase noise? Therefore increasing pitch will decrease signal noise ratio, leading to poorer contrast resolution

    • @will_creene
      @will_creene Před rokem

      Hi Diana, thanks for the question. Signal to noise ratio is essentially a function of how much information you are collecting. Anything that decreases your radiation dose from the source is going to decrease the total amount of radiation hitting the detector, and decrease our overall signal. Less signal means more noise. Regarding pitch - if the pitch is higher, that means the patient is moving through the scanner faster relative to the speed of gantry rotation. There is less exposure time per volume of body tissue, and for pitch >1, there "gaps" in between helical rotation in which information is not acquired from all sides at all levels in the Z-axis. All of this combines to mean less radiation dose to the patient, less radiation to the detector, less signal, and thus lower signal to noise ratio (= more noise!) Hope this makes sense :)

  • @growfuture5532
    @growfuture5532 Před rokem

    Thanku sir for this session

  • @nanda6027
    @nanda6027 Před rokem

    Play back speed is too fast 🤷‍♂️

  • @Rj-rs4dn
    @Rj-rs4dn Před rokem

    Hello sir what are your sources?

  • @Mesh43
    @Mesh43 Před rokem

    Thanks!

  • @dem890
    @dem890 Před rokem

    Thank you so much

  • @lindemann316
    @lindemann316 Před rokem

    Sir, your channel is pure gold, so much information condensed in a small video. Good job!

  • @aggelikikonstanti1902

    this was such a good explanation...you have single-handedly saved my grade....

  • @is6616
    @is6616 Před rokem

    This is a great video, well explained and very easy to follow! Thank you so much:)

  • @_believer_07
    @_believer_07 Před rokem

    Beautiful Explanation Ever 💞

  • @pinkpickledcauliflower9

    Absolutely *love* this. nice job.

  • @thecreator1127
    @thecreator1127 Před rokem

    Very useful Thank You So Much

  • @thecreator1127
    @thecreator1127 Před rokem

    Thank You So Much

  • @thecreator1127
    @thecreator1127 Před rokem

    Thank you so much

  • @theabhizing
    @theabhizing Před rokem

    Thanku 😍❤️

  • @wavplyr1
    @wavplyr1 Před rokem

    WOW!!!! Thank you!!! Very informative and knowledgeable!!!!

  • @sarahmccoy1941
    @sarahmccoy1941 Před rokem

    Lots of radiation. Does not sound good, whatever the equation.

  • @jkaisimm
    @jkaisimm Před rokem

    Solid. Thanks for the video! Very helpful as I read a chapter on head and neck CT imaging!

  • @rongmoshal614
    @rongmoshal614 Před rokem

    Very good explanation

  • @tarasinghSharif
    @tarasinghSharif Před rokem

    Do u have same explanation videos for Petct?